Commissioning a multileaf collimator virtual cone for the stereotactic radiosurgery of trigeminal neuralgia

Thomas A. D. Brown,1 Rex G. Ayers,1 Richard A. Popple2
1Northwest Medical Physics Center, Lynnwood, WA; 2University of Alabama at Birmingham, Birmingham, AL

Journal of Applied Clinical Medical Physics, 23, 4 (2022).

A multileaf collimator (MLC), virtual-cone treatment technique has been commissioned for trigeminal neuralgia (TGN) at Tri-Cities Cancer Center (TCCC). This novel technique was initially developed at the University of Alabama in Birmingham (UAB); it is designed to produce a spherical dose profile similar to a fixed, 5-mm conical collimator distribution. Treatment is delivered with a 10-MV flattening-filter-free (FFF) beam using a high-definition MLC on a Varian Edge linear accelerator. Absolute dose output and profile measurements were performed in a 20 × 20 × 14 cm3 solid-water phantom using an Exradin W2 scintillation detector and Gafchromic EBT3 film. Dose output constancy for the virtual cone was evaluated over 6 months using an Exradin A11 parallel plate chamber. The photo-neutron dose generated by these treatments was assessed at distances of 50 and 100 cm from isocenter using a Ludlum Model 30–7 Series Neutron Meter. TGN treatments at TCCC have been previously delivered at 6-MV FFF using a 5-mm stereotactic cone. To assess the dosimetric impact of using a virtual cone, eight patients previously treated for TGN with a 5-mm cone were re-planned using a virtual cone. Seven patients have now been treated for TGN using a virtual cone at TCCC. Patient-specific quality assurance was performed for each patient using Gafchromic EBT-XD film inside a Standard Imaging Stereotactic Dose Verification Phantom. The commissioning results demonstrate that the virtual-cone dosimetry, first described at UAB, is reproducible on a second Edge linear accelerator at an independent clinical site. The virtual cone is a credible alternative to a physical, stereotactic cone for the treatment of TGN at TCCC.

Post navigation

Previous Previous
Medical physics as an anchor for physics learning
NextContinue
Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom
Northwest Medical Physics Center

Garden Grove Office Park
11314 4th Ave W
Suite 110
Everett, WA 98204

Phone: (425) 672-2841

Linkedin Linkedin
  • Services
  • About Us
  • Contact Us
  • Privacy Policy

Physicists' Corner

Remote Web Access

SharePoint

Webmail

© 2026 Northwest Medical Physics Center

  • Services
    • Our Services
    • Staffing
    • Shielding
    • Annuals
    • Commissioning
    • Diagnostic & Health Physics Services
    • Treatment Technologies
    • Service Request Form
  • About
    • About Us
    • Our Mission
    • Our Team
    • Residency & Education
    • Publications
  • Careers
  • Contact Us
  • Physicists’ Corner
  • Job Opening – Medical Physicist – Anchorage, AK